What is cyclic voltammetry (CV)? How is a voltammogram read?
1 Theory
Excitation and response
In CV the potential is scanned as a triangular wave between two limits at constant rate (v, V/s) and the current is recorded. For a reversible couple, during the forward scan the species at the surface is depleted while the diffusion layer grows; the current first rises, then decreases under transport limitation — this is how the peak forms. On the reverse scan the product is converted back and a peak in the opposite direction is seen.
Diagnostic criteria
In a reversible (Nernstian) system: ΔE_p = E_pa − E_pc ≈ 59/n mV (25 °C), peak-current ratio i_pa/i_pc ≈ 1, peak potentials independent of scan rate and peak current proportional to v^(1/2) according to the Randles–Ševčík equation: i_p = 2.69×10⁵·n^(3/2)·A·D^(1/2)·c·v^(1/2). In a quasi-reversible system ΔE_p grows with scan rate (k⁰ is calculated by the Nicholson method); in an irreversible system the reverse peak disappears. For surface-attached (adsorbed) species the peak current is directly proportional to v and ideally ΔE_p = 0 — this criterion separates a diffusion-controlled process from a surface-controlled one.
Mechanism diagnosis
Coupled chemical reactions (EC, ECE, catalytic EC′) change the peak ratios and potential shifts in characteristic ways; a scan-rate series is the basic tool for distinguishing these mechanisms. Overlaying experimental curves with digital simulation (finite differences) allows quantitative determination of k⁰, α and chemical rate constants.
Practical settings
Scan rates range from 1 mV/s (solid state, batteries) to 100 V/s+ (microelectrodes, fast kinetics); at high rates the capacitive current (i_C ∝ v) swamps the Faradaic current (∝ v^(1/2)) and the signal-to-background ratio falls. Potential step (resolution), filters and ohmic-drop compensation directly affect the peak shape. On battery electrodes CV is a fast map of redox potentials and phase transformations; on supercapacitors a near-rectangular curve indicates ideal capacitive behaviour.
2 Application in measurement
In CV the current is recorded while the potential is scanned back and forth at constant rate. For a reversible redox couple the peak separation is about 57/n mV and the peak current is proportional to the square root of the scan rate (Randles-Ševčík); for adsorbed species the peak current is directly proportional to the rate. Peak potentials, reversibility, electron number and accompanying chemical steps (EC, ECE) are extracted from the voltammogram.
Practical rules: record the potential window and the blank voltammogram of the supporting electrolyte first; choose the scan rate so that it does not increase ohmic drop; measure at different rates to separate capacitive from Faradaic current. An internal standard such as ferrocene corrects reference drift in non-aqueous media.
The EC-Lab® CV Sim/CV Fit tools let you test a mechanism hypothesis by simulation; the Entek potentiostat, with its built-in sweep generator, implements CV in teaching laboratories with a recorder/PC.
Key equations
Randles-Ševčík (25 °C): i_p = 2.69×10⁵·n^(3/2)·A·D^(1/2)·C·v^(1/2)
3 Applicability
Where it applies
- The first 'fingerprint' of a new system: redox potentials, reversibility, electrolyte window.
- Catalyst ECSA determination, electropolymerisation, phase transformation and SEI formation on battery electrodes.
- Kinetic parameter determination (scan-rate series + simulation).
Limits and caveats
- Low sensitivity for trace-concentration analysis; pulsed techniques are preferred.
- In systems with high ohmic drop the kinetic interpretation can be misleading; R_u must be corrected first.
4 What did the application notes find?
- Kinetics AN41-1 — The effect of scan rate on peak potential and current in a simple E mechanism is simulated with CV Sim; the transition from reversible to irreversible as k⁰ decreases is followed numerically.
- Kinetics AN41-2 — The effect of ohmic drop and double-layer capacitance on the CV shape is shown by simulation: R_u enlarges the peak separation, C_dl shifts the curve with a tilted background; together they can mimic slow kinetics.
- Battery AN26 — Shows how multi-cycle CV and battery data are plotted with customisation in the software and how derived quantities (e.g. capacity, dQ/dV) are generated with the calculator.